Complete Operation Tutorial for the easyEndo™ Lite Series Surgical Stapler

This tutorial provides a detailed, step-by-step guide to the proper use of the easyEndo™ Lite series surgical stapler. Follow each section carefully to ensure optimal performance and patient safety.




Table of Contents


1. Instrument Preparation

2. Loading the Cartridge

3. Insertion of the Instrument

4. Firing Process

5. Releasing the Instrument

6. Removal and Unloading

7. Precautions and Final Checks




Instrument Preparation


Before starting any surgical procedure, it is crucial to prepare your instrument correctly.


Unpack the Device:

1. Open the graphic carton and remove the instruction manual along with the instrument, which is securely placed in its plastic packaging.

2. Follow the arrow on the dialysis paper to tear it off, then extract the stapler.

3. Remove the single-use loading unit in a similar manner.


Unpack the Device


Pre-Loading Checks:

1. Trigger: Ensure the trigger moves smoothly.

2. Stapler: Clear the stapler of any obstructions.

3. Rotating Knob: Confirm that the rotating knob is in the middle position.

4. Yellow Protective Cap: Do not remove this cap until the cartridge is loaded.

5. Cartridge Inspection: Verify that the cartridge’s yellow protective cap is correctly positioned and ensure the jaw remains open during installation.


Pre-Loading Checks

Pre-Loading Checks

Pre-Loading Checks




Loading the Cartridge


Proper cartridge loading is essential for the device's functionality.


1. Insert the Cartridge:

Slide the cartridge from the bottom into the cartridge.

Align it with the alignment tab, and press firmly until an audible “click” confirms proper placement.


image.png


2. Remove the Protective Cap:

Once the cartridge is securely in place, remove the yellow protective cap.


Remove the Protective Cap




Insertion of the Instrument


Correct insertion is critical for safe and effective surgical performance.


1. Activate the Trigger:

Pull the trigger to close the jaw, ensuring that it makes full contact with the handle.


Activate the Trigger


2. Trocar Insertion:

Insert the stapler into the patient’s body via a suitable trocar.

Confirm that the double articulating joints clear the trocar completely.


Trocar Insertion


3. Adjust the Angle:

To open the jaws, either press the release button or pull back the release knob.

Rotate the wrench to adjust the angle to a maximum of 60°. The instrument offers four preset positions: 15°, 30°, 45°, and 60° (with independent gear adjustments for each side).


Adjust the Angle




Firing Process


The firing process secures the tissue and completes the stapling sequence.


1. Clamp the Tissue:

Ensure that the target tissue is clamped securely.

Clamp the Tissue


2. Engage the Trigger:

Press the trigger firmly, maintaining a tight grip on the tissue.


3. Timed Delay:

Wait for 15 seconds after clamping before proceeding.

Timed Delay


4. Safety and Firing:

Pull the green safety.

Press the trigger repeatedly in controlled steps until the entire firing process is complete.


5. Completion Check:

Confirm the process is finished by observing the cutting indicator under the microscope or by monitoring the release button on the stapler.


Completion Check




Releasing the Instrument


After the firing process, safely release the instrument.


1. Open the Jaws:

Pull back the release button fully to open the jaw.

Open the Jaws


2. Reset Articulating Cartridge:

For articulating cartridges, turn the rotating wrench to return the cartridge to its original position.


Reset Articulating Cartridge


3. Check for Bleeding:

Ensure there is no bleeding on the sutured tissue before proceeding.


Check for Bleeding




Removal and Unloading


Final steps involve safely removing the instrument and unloading the used cartridge.


1. Extract the Stapler:

Close the jaw and carefully remove the stapler through the trocar.


2. Unload the Cartridge:

Push the distal end of the cartridge to detach it from the stapler.


Unload the Cartridge




Precautions and Final Checks


  • Pre-Use Inspection: Always verify that the trigger and stapler mechanism are smooth and clear.

  • Cartridge Installation: Ensure the yellow protective cap remains in place until the cartridge is loaded.

  • Safety First: Follow each step meticulously to avoid any complications during surgery.


Following these instructions will ensure that the easyEndo™ Lite series stapler is used safely and effectively, promoting better surgical outcomes. For further information or troubleshooting, refer to the detailed user manual provided in the packaging or contact EziSurg Medical support.




This comprehensive tutorial is designed to assist medical professionals in the effective use of the easyEndo™ Lite series surgical stapler, ensuring precision and safety in every procedure.


Related Information

Related Keywords

Surgical Stapling ProductsWhy Disposable Laparoscopic Instruments Are Reshaping Operating RoomsPowered Staplers for Gastrointestinal Surgery|EziSurg Medical45mm vs 60mm Endoscopic Stapler: How to Choose Length | EziSurg Medical5 mm vs 10/12 mm Trocars: How to Choose Port Size for Laparoscopic ProceduresChoosing the right surgical stapler for your operating roomEnhancing Surgical Precision with Advanced Staplers and EndoscopesEssential Tools Used in Laparoscopic Surgery: A Comprehensive GuideHow EziSurg Designs the Best Minimally Invasive Surgical Instruments for LaparoscopyWhy Choose Disposable Surgical Instruments for Optimal Patient Safety?Understanding Bariatric Surgery | EziSurg MedicalEnhancing Outcomes in Colorectal Surgery: EziSurg Medical Staple TechnologyPerforming Tissue Dissection Procedures with Energy Sealing SystemsSingle-Use Surgical Instruments Can Be a Cost-Effective Choice for HospitalsAnalysis of Clinical Outcomes Using Energy Sealing SystemsGet to know linear stapler surgery better with EziSurgFrom Cutting to Closure: Complete Anastomosis Solutions for SurgeonstestddLaparoscopic Surgical Instruments from Ezisurg MedicalWhat is Ultrasonic Surgical System?A Comprehensive Guide to the Surgical Stapler Structure in Stapler SurgerySurgical Staple Gun: Understanding the Device and Its Evolving Role in SurgeryMinimally Invasive Surgery Solutions from EziSurg MedicalSurgical Stapling Devices For Sale| The cheapest priceAnalysis of Clinical Outcomes Using EziSurg's Minimally Invasive Laparoscopic InstrumentsColumn headingsArticulation Explained: When 45°/60°/120° Matters in MISLinear Stapler SurgicalSurgical Stapling DevicesDisposable vs. Reusable Laparoscopic Instruments: Pros and Cons for HospitalsEndoscopic Linear Cutter StaplersAdvanced Surgical Stapling Solutions for Colorectal SurgeryIndications and Key Applications of the easyEndo™ Lite Surgical StaplerEziSurg Collaboration with Top Doctors: Elevating Surgical ExcellenceBenefits of Single-Use Flexible Ureteroscopes in Modern UrologyThe Role of EziSurg Medical Surgical Staplers in Thoracic SurgeryEziSurg Medical Staplers: Secure Stapling for Bariatric SurgeryIntestinal Anastomosis StaplerLaparoscopic StaplerHow Does a Vessel Sealing Device Work? Different Types of Vessel Sealing MachinesPowered vs. Manual Surgical Staplers: Which to Choose?Suture vs. Staple: When to Use Each in Surgical ProceduresEziSurg surgical innovationsSurgical Stapling Solutions: Advanced Tissue Closure TechnologyEndo Linear Cutter Stapler

Headquarters (China)

Address

No. 399 Miaoqiao Road, Pudong New District, Shanghai 201315, P.R. China

Phone

+86 400-090-6176

Fax

+86 21-5067-6156

Website

https://en.ezisurg.com

U.S. Subsidiary

Address

Suite 403, 6136 Frisco Square Blvd., Frisco, Texas 75034, USA

Phone

+1 469-252-2118

Or Leave Us a Message, We Will Get Back to You Asap!

Copyright © Ezisurg Medical Co., Ltd. All Rights Reserved. | Sitemap Privacy Policy Terms of Services | Powered by Reanod

WeChat